Wavelength range UV-Vis-NIR ; Broadband high throughput achromatic mirror based system, optimized from 300 nm to 1600 nm without changing optics. Standard laser wavelengths 532, 785 nm.
Typical lasers for Raman and PL Spectrometer scanning speed Up to 400 nm/s with 600g/mm grating, mounted on a standard Turbodrive 4-grating turret, for fast Raman, PL spectral map acquisitions. Number of gratings unlimited 4 grating exchangeable motorized turret.
Fast Imaging <1ms/spectrum SWIFT, SWIFT XS EMCCD, SWIFT repetitive, SWIFT extended Range and SmartSampling for ultrafast imaging. 3D confocal slicing with laser lightsheet illumination.
QScan (patented) Typical 2 µm thick 100 x 100 µm2 laser lightsheets for large area 2D/3D confocal imaging.
Standard wavenumber cut-off 30 cm-1 With edge filters for 532, 638 and 785 nm wavelengths, injection rejection, > 99% transmission
Laser auto-alignment procedure 15 s
Lasers Up to 4 built-in solid-state lasers + 1 external laser NUV to NIR wavelengths available.
Spectroscopic motorized modalities Up to 6 laser filters Raman, PL, Super Low Frequency …
Rayleigh filter orientation Individual filter computer controlled
Factory preset & user adjustable angle to adapt with sample reflectivity